Comparison study of sound touch quantification and acoustic radiation force impulse imaging in measuring liver and spleen stiffness

Qingjuan Wang1, Jian Zheng1,(), Rizhen Gu1, Xiaohong Yao1, Siliang Kang1, Lexiang Long1, Dan Zhou1, Xiaoli Wang2, Rongqin Zheng2   

  1. 1. Department of Ultrasound, the Third People’s Hospital ofLonggang District, Shenzhen 518115, China
    2. Department of Ultrasound, the Third Affiliated Hospital of Sun Yat-sen University, Guangzhou 510630, China
  • Received:2019-06-01 Online:2019-10-01 Published:2019-10-01
  • Contact: Jian Zheng
  • About author:
    Corresponding author: Zheng Jian, Email:

Abstract:

Objective

To compare sound touch quantification (STQ) and acoustic radiation force impulse imaging (ARFI) in measuring liver and spleen stiffness.

Methods

From January 2018 to April 2019, using STQ and ARFI technology, a total of one hundred and seventy subjects (71 healthy volunteers and 99 patients with chronic hepatitis B) were tested for liver and spleen stiffness. The success rate, variability and measured values of the two techniques were compared.

Results

The success rate of liver STQ and ARFI was 96.5% (164/170) and 95.9% (163/170) (χ2<0.001, P=0.996), and the variability was (6.48±4.72) %and (8.75±5.01) % (t=4.127, P<0.001), respectively. The success rate of spleen STQ and ARFI was 94.7% (161/170) and 92.9% (158/170) (χ2=0.215, P=0.643), and the variability was (7.34±4.26) % and (11.04±4.77) % (t=6.282, P<0.001), respectively. Normal liver STQ and ARFI values were (1.45±0.18) m/sand (1.15±0.30) m/s (t=7.565, P<0.001), and chronic hepatitis B patients were (1.59±0.26) m/s and (1.24±0.30) m/s (t=14.005, P<0.001). The normal spleen STQ and ARFI values were (2.31±0.25) m/sand (1.96±0.45) m/s (t=5.555, P<0.001), and chronic hepatitis B patients were (2.33±0.25) m/s and (2.08±0.42) m/s (t=5.396, P<0.001), respectively.

Conclusions

The success rate of STQ and ARFI is high and the variability is low. Compared with ARFI, STQ further improves the stability of measurement. In addition, the measured values of STQ and ARFI cannot be directly equivalent, which needs to be distinguished in clinical application.